FCMPASS Detector Optimisation: Data Output


Upon completion of detector incrementation analysis the optimal gains for small particle analysis will be shown. If the rainbow beads were cross-calibrated the sensitivity of the lower limit of detection will also be shown.

On the Cytek Aurora platform, an admin template settings file with the optimal gains will also be exported. This template will not normalize to QC data unless alterations are made and saved.

Experiment calibration

An unnormalized QC plot is outputted to demonstrate the linearity of the system. Peaks that are not resolved from the noise population are removed for illustrative purposes.

Experiment calibration

A normalized QC plot is outputted to demonstrate the detector settings with the highest signal-to-noise. Peaks that are not resolved from the noise population are removed for illustrative purposes.

Experiment calibration