FCMPASS Detector Optimisation: Data Generation
An up-to-date protocol for the acquiring of 8-peak detector setting incremenetation data can be found in the protocol below.
Auto-generating Cytek Aurora Gain Templates
- To increase the ergonomics of running detector setting incrementation on the Cytek Aurora, FCMPASS can auto-generate fixed detector template settings that will not normalize to QC settings unless the templates are modified and saved.

- Navigate to the detector optimzation module from the start menu. Click 'Menu' and select 'Voltration Template Generator'.

- Modify the gains and threshold accordingly (use the protocol above to aid the selection of settings) and click 'Generate Template'.
- Select a folder to exported the generated templates to.
The settings listed are a guide and may be sub-optimal for your specific instrument.